Hydrometer temperature correction
A hydrometer calibrated at 60 °F reads low in hot syrup. Correct a reading for the syrup’s temperature, or find the reading that means the syrup is on target.
01 · Inputs
Pick what you want to know and the scale on your hydrometer, then enter the syrup temperature at the moment you read it.
I want to
Hydrometer scale
Read where the syrup surface crosses the scale, not the top of the meniscus.
Measure it right before you read the hydrometer; hot syrup cools fast. The table covers 40°F to 212°F.
Hydrometer card
What a hydrometer calibrated at 60°F should read in syrup that is exactly 66.0 or 66.9 °Brix, on the scale chosen above. Print it and keep it by the draw-off.
| Syrup at | Correction | Reads for 66.0°Brix | Reads for 66.9°Brix |
|---|---|---|---|
| 60°F | 0.00°Brix | 66.0°Brix | 66.9°Brix |
| 70°F | +0.50°Brix | 65.5°Brix | 66.4°Brix |
| 80°F | +1.00°Brix | 65.0°Brix | 65.9°Brix |
| 90°F | +1.50°Brix | 64.5°Brix | 65.4°Brix |
| 100°F | +2.00°Brix | 64.0°Brix | 64.9°Brix |
| 110°F | +2.50°Brix | 63.5°Brix | 64.4°Brix |
| 120°F | +3.00°Brix | 63.0°Brix | 63.9°Brix |
| 130°F | +3.50°Brix | 62.5°Brix | 63.4°Brix |
| 140°F | +4.00°Brix | 62.0°Brix | 62.9°Brix |
| 150°F | +4.56°Brix | 61.4°Brix | 62.3°Brix |
| 160°F | +5.11°Brix | 60.9°Brix | 61.8°Brix |
| 170°F | +5.67°Brix | 60.3°Brix | 61.2°Brix |
| 180°F | +6.22°Brix | 59.8°Brix | 60.7°Brix |
| 190°F * | +6.81°Brix | 59.2°Brix | 60.1°Brix |
| 200°F * | +7.39°Brix | 58.6°Brix | 59.5°Brix |
| 211°F *(hot test) | +8.00°Brix | 58.0°Brix | 58.9°Brix |
* Above 180°F, extrapolated in the manual’s table rather than measured. Confirm hot readings after the syrup cools.
03How this is calculated
Most maple hydrometers are calibrated at 60 °F (15.6 °C). Hot syrup is less dense, so the hydrometer sinks deeper and reads low. The calculator adds the correction from Table 8.2 of the North American Maple Syrup Producers Manual for the syrup’s temperature, interpolating between rows: +1.0 °Brix at 80 °F, +4.0 at 140 °F, +8.0 from 209 °F up. A Baumé reading takes the table’s Baumé column, which is half the Brix one. The manual’s example: 65.8 °Brix read at 80 °F is 66.8 °Brix.
Working backward gives the reading that means the syrup is on target, the same idea as a Murphy Cup, whose dial shows the reading for 66.9 °Brix at the cup’s temperature. At 211 °F it is the hot test line: the target less 8 °Brix (4 °Baumé), so 58.9 °Brix or 32 °Baumé for 66.9. Corrected values are converted between Brix and Baumé with the National Bureau of Standards density tables: 66.9 °Brix is 36.0 °Baumé, the hydrometer’s red line, and 66.0 is about 35.5, where the manual prints 35.6.
The manual is clear that hot readings, particularly above 180 °F, are relatively inaccurate: the corrections there are extrapolated, not measured, and the syrup cools while you read it. It recommends confirming hot readings once the syrup is near room temperature, where the hydrometer is most accurate.
A hydrometer is only as good as its glass and paper. Vermont law requires hydrometers used for maple syrup in the state to be tested and marked by the Vermont Agency of Agriculture, Food & Markets. The manual recommends checking every hydrometer’s accuracy each year and keeping at least three to compare. Most maple hydrometers are marked for 66.9 °Brix, so where 66.0 is the minimum, finishing right on the red line gives away sugar.
Sources
- Ch. 8, "Syrup Density Scales—Brix and Baume" and Table 8.2 hydrometer temperature correctionsNAMSPM 3rd ed. (2022)p. 8-13, Table 8.2
- Ch. 8, using a hydrometer: Cold Test 60 °F and Hot Test 211 °F lines for 66.9 °Brix (36 °Bé), hot test as the desired density less 8 °Brix (4 °Bé), Vermont hydrometer testing by VAAFM, annual accuracy checkNAMSPM 3rd ed. (2022)p. 8-14
- Ch. 8, hydrometer suggestions: readings above 180 °F are relatively inaccurate and their corrections extrapolated; the Murphy Cup shows the reading a hydrometer should give at the syrup’s temperatureNAMSPM 3rd ed. (2022)pp. 8-15 to 8-16
- Ch. 8, pure maple syrup definition and legal density (66.0 to 68.9 percent soluble solids; 66.9 °Brix minimum in Vermont and New Hampshire, footnote 1)NAMSPM 3rd ed. (2022)p. 8-2
- Weights per United States Gallon and Weights per Cubic Foot of Sugar SolutionsSnyder, C.F., Hammond, L.D. (1946). National Bureau of Standards Circular C457Table 1, pp. 3–34: true specific gravity 20 °C/20 °C and pounds per US gallon in air at 20 °C, by tenth degree Brix; Table 2 for water